ResearchSpace

The IFIN-HH triple coincidence liquid scintillation counter

Show simple item record

dc.contributor.author Razdolescu, AC
dc.contributor.author Broda, R
dc.contributor.author Cassette, P
dc.contributor.author Simpson, BRS
dc.contributor.author Van Wyngaardt, WM
dc.date.accessioned 2007-06-29T08:52:57Z
dc.date.available 2007-06-29T08:52:57Z
dc.date.issued 2006-10
dc.identifier.citation Razdolescu, AC, et al. 2006. The IFIN-HH triple coincidence liquid scintillation counter. Applied Radiation and Isotopes, vol. 64(10), pp 1510-1514 en
dc.identifier.issn 0969-8043
dc.identifier.uri http://hdl.handle.net/10204/784
dc.description Available from: http://www.sciencedirect.com en
dc.description.abstract The paper summarizes the IFIN-HH triple coincidence liquid scintillation counter used for the implementation of the TDCR method. The electronic unit was recently extended to record the three individual double coincidence ratios to take into account the differences in the quantum efficiencies of the three-photomultiplier tubes. Some details of the electronic system and the data processing are given. The critical point of a TDCR counter is to adjust correctly the discriminator levels on the three channels under the single electron peak. The paper describes the method of adjustment based on the evolution of the dark counting rate versus the discriminator level. Also indicated is the influence of the discrimination level on the activity results as measured at IFIN-HH using a 3 H standard. The performances of the IFIN-HH TDCR counter was checked against the measurement results of the TDCR counters of CSIR NML (South Africa), RC (Poland) and LNHB (France). A set of ready-to-measure Ni-63 sources in liquid scintillator, in sealed counting vials, was prepared and dispatched for measurement to all these laboratories. The paper describes designs of the TDCR counters used. An analysis and discussion of the measurement results is given. en
dc.language.iso en en
dc.publisher Pergamon-Elsevier Science Ltd en
dc.subject TDCR method en
dc.subject Discrimination level en
dc.subject Liquid scintillation counter en
dc.subject IFIN-HH triple coincidence liquid scintillation en
dc.title The IFIN-HH triple coincidence liquid scintillation counter en
dc.type Article en
dc.identifier.apacitation Razdolescu, A., Broda, R., Cassette, P., Simpson, B., & Van Wyngaardt, W. (2006). The IFIN-HH triple coincidence liquid scintillation counter. http://hdl.handle.net/10204/784 en_ZA
dc.identifier.chicagocitation Razdolescu, AC, R Broda, P Cassette, BRS Simpson, and WM Van Wyngaardt "The IFIN-HH triple coincidence liquid scintillation counter." (2006) http://hdl.handle.net/10204/784 en_ZA
dc.identifier.vancouvercitation Razdolescu A, Broda R, Cassette P, Simpson B, Van Wyngaardt W. The IFIN-HH triple coincidence liquid scintillation counter. 2006; http://hdl.handle.net/10204/784. en_ZA
dc.identifier.ris TY - Article AU - Razdolescu, AC AU - Broda, R AU - Cassette, P AU - Simpson, BRS AU - Van Wyngaardt, WM AB - The paper summarizes the IFIN-HH triple coincidence liquid scintillation counter used for the implementation of the TDCR method. The electronic unit was recently extended to record the three individual double coincidence ratios to take into account the differences in the quantum efficiencies of the three-photomultiplier tubes. Some details of the electronic system and the data processing are given. The critical point of a TDCR counter is to adjust correctly the discriminator levels on the three channels under the single electron peak. The paper describes the method of adjustment based on the evolution of the dark counting rate versus the discriminator level. Also indicated is the influence of the discrimination level on the activity results as measured at IFIN-HH using a 3 H standard. The performances of the IFIN-HH TDCR counter was checked against the measurement results of the TDCR counters of CSIR NML (South Africa), RC (Poland) and LNHB (France). A set of ready-to-measure Ni-63 sources in liquid scintillator, in sealed counting vials, was prepared and dispatched for measurement to all these laboratories. The paper describes designs of the TDCR counters used. An analysis and discussion of the measurement results is given. DA - 2006-10 DB - ResearchSpace DP - CSIR KW - TDCR method KW - Discrimination level KW - Liquid scintillation counter KW - IFIN-HH triple coincidence liquid scintillation LK - https://researchspace.csir.co.za PY - 2006 SM - 0969-8043 T1 - The IFIN-HH triple coincidence liquid scintillation counter TI - The IFIN-HH triple coincidence liquid scintillation counter UR - http://hdl.handle.net/10204/784 ER - en_ZA